Please use this identifier to cite or link to this item:
http://localhost:8081/xmlui/handle/123456789/14767
Full metadata record
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Gourishetty, Anil Kumar | - |
dc.date.accessioned | 2020-01-16T10:25:40Z | - |
dc.date.available | 2020-01-16T10:25:40Z | - |
dc.date.issued | 2016-05 | - |
dc.identifier.uri | http://localhost:8081/xmlui/handle/123456789/14767 | - |
dc.description.abstract | The aim of this project is to calculate the activity of different radioactive sources such as Cs- 137, Co-60, Na-22 and Co-57 using different scintillation detectors, namely, NaI(Tl), LaBr3(Ce) and LaCl3(Ce). The activity has been calculated using absolute method for single gamma emitter Cs-137 and triple gamma emitter Co-57 and using sum-peak method for double gamma emitters Co-60 and Na-22. GEANT4 simulation toolkit has been used for activity calculation of Co-57. Intrinsic peak efficiency of detectors has been taken from simulation results of GEANT4 toolkit to calculate the activity of sources using absolute method. Intrinsic peak efficiency has also been calculated using experimental and analytical methods and the result has been compared with the simulation result. | en_US |
dc.description.sponsorship | Indian Institute of Technology Roorkee. | en_US |
dc.language.iso | en | en_US |
dc.publisher | Department of Physics,IITR. | en_US |
dc.subject | Radioactive sources | en_US |
dc.subject | Scintillation Detectors | en_US |
dc.subject | detectors | en_US |
dc.subject | GEANT4(simulation toolkit) | en_US |
dc.title | Activity measurement using scintillation detector | en_US |
dc.type | Other | en_US |
Appears in Collections: | DOCTORAL THESES (Physics) |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
G25709-nidhi-D.pdf | 1.57 MB | Adobe PDF | View/Open |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.